A Valve Switching Ion Chromatography System With Integrated Automation Management

Valve switching applications are becoming popular in ion chromatography (IC). Example applications include matrix diversion or matrix elimination prior to analysis of trace components. A conventional ion chromatography system is usually equipped with a single injection valve, and installing additional valves for various applications such as two-dimensional (2-D) separations and sample preparation applications is cumbersome. The instrument detailed here has an integrated automation management module supporting various valve configurations and a reaction coil heater for postcolumn reaction applications. Here, the authors will discuss the utility of the system for various valve switching applications.

First, the instrumentation details of the system and the automation module will be discussed. Next examples of applications will be shown, including 2-D analyses of perchlorate in high-ionic-strength matrices and a matrix neutralization approach showing analysis of trace anions in concentrated base. With the 2-D analysis method in conjunction with suppressed conductivity detection, it is possible to detect low levels of perchlorate without the use of an expensive mass spectrometer.

System description

Figure 1 - The automation manager (AM) in the ICS-3000 detector/chromatography compartment (DC) is highlighted above. Mounted as a sliding tray for easy plumbing, the AM is versatile enough to handle difficult applications by supporting a variety of valve configurations.

The ICS-3000 Reagent-Free™ IC (RFIC™) system (Dionex Corp., Sunnyvale, CA) combines traditional instrument improvements for signal enhancement with valving and control for easy implementation of complex applications. The system consists of a pump (single or dual, SP or DP, respectively), eluent generator (EG), detector/chromatography compartment (DC), automation manager (AM), autosampler (AS), and tablet PC. The system can be configured as a single or dual system and incorporates the latest advancements in RFIC system technology (Figure 1). When configured as a single system, the additional pump from the DP can be used for miscellaneous applications such as pumping a deionized (DI) water stream for sample preparation applications.

From ordinary DI water, RFIC systems generate high-purity eluents that perform superior separations, and then neutralize the eluent back to DI water in the suppression step to facilitate sensitive detection using a conductivity detector. Using the EG with the continuously regenerated-trap column (CR-TC) minimizes baseline shift during the run, when pursuing eluent concentration step changes or gradients.

The DC compartment consists of three sections for separation, detection, and automation. The separator columns are housed in the DC in a controlled thermal environment. Capable of single or dual analysis, the DC houses conductivity and electrochemical detectors. Two detector cells are available for “plug-and-play” conductivity and electrochemical operation. The DC also contains two valves in the lower compartment for injection and sample preparation applications. All valves are on slide-out trays for easy installation and troubleshooting. An important feature within the DC is the AM, a slide-out tray that is configurable with up to two high-pressure valves, two low-pressure valves, and a postcolumn reaction heater. The AM facilitates valve switching applications.

The AM is mounted inside the DC compartment so that sample preparation steps such as preconcentration and matrix removal are in the near vicinity of the separator columns. In conventional IC systems, mounting the sample preparation valves close to the separator column is difficult and often leads to excessive band dispersion. Due to the close proximity of the AM to the separator columns, band dispersion issues are minimized.

Two-dimensional perchlorate analysis

Although there are no federal drinking water regulations for perchlorate, various states have adopted their own advisory levels that range from 1 to 18 ppb. Trace-level analysis in IC is typically done using concentrator columns or large- volume injections. When high levels of matrix ions are present, large-volume injections are preferred over concentration methods since the matrix ions can elute off the analyte of interest from the concentrator column. It should be noted that large-volume injections not only enhance the sensitivity of trace components, but also enhance the sensitivity of matrix components. In some cases, the matrix components interfere with the analysis by coeluting or eluting the trace component into a broad peak, leading to poor detection. This is the case with perchlorate analysis.

Perchlorate in drinking water is typically analyzed in the ppb range, whereas the matrix concentrations are at the hundreds-of-ppm level. U.S. EPA Method 314.0 was a direct injection method but required an off-line matrix elimination step with solid-phase extraction cartridges for samples containing high levels of matrix ions. U.S. EPA Method 314.1 was published as an update and allowed improved detection of perchlorate in high-ionic-strength matrices. Alternatively, an automated 2-D heart-cutting method can be used as an in-line approach.

The ICS-3000 system was configured as a two-channel dual system for this application. The injection valve in the first dimension was fitted with a large-volume sample loop (4 mL). The injection valve in the second dimension was fitted with a concentrator column that focused the heart-cut analyte peak from the first dimension for further analysis in the second dimension.

Figure 2 - Diagram highlighting components and configuration for 2-D IC in the ICS-3000.

The schematic for matrix removal and signal enhancement is shown in Figure 2. There are several advantages of the 2- D matrix diversion approach. Initial sample loading onto the 4-mm column allows a large sample injection volume (large amount of sample) due to the high capacity of the analytical column and higher selectivity for perchlorate relative to the matrix ions. Second, it is possible to focus the perchlorate peak that is partially resolved in the first dimension onto a concentrator column in the second dimension. The suppressed effluent with hydroxide eluent is water, which provides the ideal environment for ion-exchange retention and focusing. Third, the second dimension is operated at a lower flow rate relative to the first dimension, thereby enhancing the detection sensitivity. Finally, this approach also allows the potential to combine two different chemistries in two dimensions, thereby enabling a selectivity not possible using only a single chemistry dimension.

Figure 3 - Chromatograms of a 1-D (a) and 2-D (b) analysis of perchlorate in a high-salt matrix. Note that perchlorate is not detected due to interference from the matrix ions in the 1-D separation. In the 2-D separation, the interfering matrix has been removed and the perchlorate peak is enhanced and also quantifiable.

Figure 3a shows the analysis of a sample in the first dimension consisting of 5 ppb perchlorate in the presence of 1000 ppm matrix ions. As can be seen in this figure, the matrix interferes with the detection of the perchlorate peak, which is broadened and difficult to quantify. This same sample was analyzed using the 2-D approach by using the 2-mm column in a second dimension. As shown in Figure 3b, perchlorate is well resolved and free from any matrix effects. Chromatography with matrix diversion and concentration followed by analysis on a 2-mm column and a concentration-sensitive detector yields sensitivity proportional to the flow rate ratio of the first dimension versus the second dimension: a fourfold gain. Due to the physical proximity of the two dimensions in the system and minimal delay volume, it was possible to achieve excellent peak shape and recovery.

AutoNeutralization

Another example of valve switching involves a technique known as AutoNeutralization(Dionex Corp.). This technique is used when it is necessary to quantify anionic contaminants in concentrated bases. The strategy most often employed is to dilute the sample. This dilution reduces the concentration of the interfering matrix ion to a level that does not affect separation. However, dilution also reduces the concentration of trace anions, compromising their detection.

AutoNeutralization  solves the analytical problem of achieving good detection limits of trace anions in concentrated bases by neutralizing the base using a membrane-based neutralizer device. The sample anions are in a water background after neutralization and can be focused back onto a concentrator column that is located on a third valve in the lower compartment.

In this setup, the system was used in conjunction with an AM module.

Two six-port valves on the AM module were used for neutralization. The first valve was used for loading the sample onto an ASRN neutralizer device (Dionex Corp.). The second valve was used for holding the sample in the collection loop and rerouting the sample back through the neutralizer and diverting the neutralized stream onto a concentrator column.

Figure 4 - Plumbing configuration required for AutoNeutralization applications.

This plumbing configuration is shown schematically in Figure 4. In the AutoNeutralization process, the concentrated base sample is loaded into the 25-μL sample loop of the sample valve. The sample loop is switched in-line and flushed with a stream of deionized water (also known as the carrier solution). The carrier solution transfers the concentrated base from the sample loop to the ASRN, where the sample is partially neutralized, then transferred to the 5000-μL loop. The recycle valve is then actuated to pass the sample through the ASRN again so that it is completely neutralized. The completely neutralized sample is finally delivered to a concentrator column; since the anions are now in water, they are concentrated as a tight band. Finally, the anions are eluted from the trap column to the analytical column, separated, and detected.

Figure 5 - Chromatographic results of anion analysis with (a) and without (b) AutoNeutralization, from caustic sample matrix.

Figure 5 shows the results of anion analysis of a concentrated base sample with (a) and without (b) neutralization. Without neutralization, the anions are broadened and show poor sensitivity in a 1% sodium hydroxide sample matrix. In the 50% sample matrix the presence of the matrix causes a high background that makes detection of additional anions virtually impossible. Figure 5a shows the results after neutralization. Note that the large background has been removed and detection of anions is now possible with good recovery.

The AM allows easy implementation of the autoneutralization application without any additional external valves or controls. The system configuration for this application was simplified with the AM installed. All valves, plumbing, trap columns, and ASRN remain inside the DC compartment, minimizing tubing lengths and allowing all manipulation of sample to be contained in a precisely controlled environment. In addition, all valves are recognized through Chromeleon® control software (Dionex Corp.), eliminating the use of external TTL controls or triggers.

Conclusion

The ICS-3000 system provides an effective platform for pursuing standard and multiple valve switching applications. Applications have been shown that demonstrate the utility of the system and the AM module for valve switching applications. Due to the close proximity of the valves to the columns and detector cells, this design provides a low-dispersion platform for implementing multidimensional sample preparation and analysis schemes. The inherent flexibility and precise control of the system make such nonroutine applications easier to implement.

Dr. Jack, Dr. Lin, Mr. De Borba, and Dr. Srinivasan are with Dionex Corp., 1228 Titan Way, Sunnyvale, CA 94086-4015, U.S.A.; tel.: 408-737-0700; fax: 408-739-4398; e-mail: [email protected]. Dr. Sekiguchi, Mr. Nakanishi, and Ms. Yoshimura are with Nippon Dionex, K.K., Osaka, Japan.